Assessment of CareStart G6PD rapid diagnostic test and CareStart G6PD biosensor in Mauritania

Abstract

Background

The elimination of Plasmodium vivax malaria requires 8-aminoquinolines, which are contraindicated in patients with glucose-6-phosphate dehydrogenase (G6PD) deficiency due to the risk of acute haemolytic anaemia. Several point-of-care devices have been developed to detect G6PD deficiency. The objective of the present study was to evaluate the performance of two of these devices against G6PD genotypes in Mauritania.

Methods

Outpatients were screened for G6PD deficiency using CareStart™ rapid diagnostic test (RDT) and CareStart™ G6PD biosensor in Nouakchott, Mauritania, in 2019–2020. African-type and Mediterranean-type G6PD genotypes commonly observed in Africa were determined by polymerase chain reaction-restriction fragment length polymorphism and sequencing. Qualitative variables were compared using Fisher’s exact test.

Results

Of 323 patients (74 males and 249 females), 5 males and 2 homozygous females had the African-type A- genotype: A−(202) in 3 males and 2 females and G6PD A−(968) in 2 males. Among heterozygous females, 13 carried G6PD A−(202), 12 G6PD A−(968), and 3 G6PD A−(542) variants. None had the Mediterranean-type G6PD genotype. Eight had a positive G6PD RDT result, including all 7 hemizygous males and homozygous females with A- or A-A- (0.12 to 2.34 IU/g haemoglobin, according to G6PD biosensor), but RDT performed poorly (sensitivity, 11.1% at the cut-off level of < 30%) and yielded many false negative tests. Thirty-seven (50.0%) males and 141 (56.6%) females were anaemic. The adjusted median values of G6PD activity were 5.72 and 5.34 IU/g haemoglobin in non-anaemic males (n = 35) and non-anaemic males and females (n = 130) with normal G6PD genotypes using G6PD biosensor, respectively. Based on the adjusted median of 5.34 IU/g haemoglobin, the performance of G6PD biosensor against genotyping was as follows: at 30% cut-off, the sensitivity and specificity were 85.7% and 91.7%, respectively, and at 80% cut-off, the sensitivity was 100% while the specificity was 64.9%.

Conclusions

Although this pilot study supports the utility of biosensor to screen for G6PD deficiency in patients, further investigation in parallel with spectrophotometry is required to promote and validate a more extensive use of this point-of-care device in areas where P. vivax is highly prevalent in Mauritania.

Graphic abstract

Background

Glucose-6-phosphate dehydrogenase (G6PD) is the key enzyme of the oxidative phosphogluconate pathway, the only source of nicotinamide adenine dinucleotide phosphate hydrogen (NADPH) in mature erythrocytes, required to generate glutathione and protect erythrocytes from oxidative damage. Human G6PD is encoded by a housekeeping X-linked gene spanning 18.5 kilobases, which is composed of 13 exons and 12 introns [1, 2]. The G6PD gene has undergone numerous mutations, leading to more than 200 G6PD variants with a wide range of enzyme activity [3,4,5]. Some variants lead to clinical signs and symptoms, but the majority remain clinically silent, and blood profiles are normal. However, when G6PD-deficient individuals are exposed to exogenous triggering factors (drugs, fava beans, infections), low levels of intraerythrocytic NADPH and rapid depletion of glutathione may result in accumulation of reactive oxygen species and destruction of erythrocytes (haemolysis), haemoglobinuria, and hyperbilirubinaemia [6].

G6PD deficiency occurs frequently in human populations residing in malaria endemic areas around the world [7]. Acute haemolytic anaemia has been reported in G6PD-deficient individuals after administration of certain antimalarial drugs, including dapsone, methylene blue, and 8-aminoquinolines (primaquine and tafenoquine) [8,9,10,11,12,13,14]. The degree of dapsone- and primaquine-induced acute haemolytic anaemia is dose-dependent [9, 15]. Despite this well-known, potentially fatal adverse effect, 8-aminoquinolines are indispensable to achieve elimination of both Plasmodium falciparum and Plasmodium vivax [16]. The 8-aminoquinolines have distinct roles for each of these Plasmodium species. Mature P. falciparum gametocytes survive exposure to blood schizonticides but can be killed with a single low-dose primaquine, which can be administered without prior testing of G6PD status [17,18,19]. The complete treatment of P. vivax malaria, referred to as “radical treatment,” consists of administering at least two drugs targeted against different asexual developmental stages of the parasite in the human host [16]. A blood schizonticide (chloroquine or any artemisinin-based combination therapy, except the combination with sulfadoxine-pyrimethamine) is given to kill the asexual blood stage rapidly to relieve the patient of acute signs and symptoms. A second drug, an 8-aminoquinoline, is given either daily (for primaquine) over 14 days or in a single dose (for tafenoquine) to kill the hypnozoites lying dormant in the hepatocytes and prevent relapse [16, 20,21,22]. The anti-hypnozoite therapy with either primaquine or tafenoquine is effective, safe, and relatively well-tolerated in a compliant patient with a normal G6PD activity. However, due to the major drawback of 8-aminoquinolines, they are contraindicated in individuals with G6PD deficiency [16].

Mauritania is the only country in West Africa where P. vivax has become highly endemic in recent years [23]. Since 2013, the Mauritanian Ministry of Health recommends the systematic use of primaquine, in addition to artesunate-amodiaquine or artemether-lumefantrine, to treat all cases of P. vivax malaria, in line with the World Health Organization (WHO) recommendations [16, 24, 25]. However, to date, the new drug policy has remained a dead letter, not only because primaquine (or tafenoquine) is not available in the country, but also because health care workers do not have access to information about the G6PD status of individual patients, which requires the measurement of G6PD enzyme activity using ultraviolet (UV) spectrophotometry as the gold standard method [26,27,28,29].

To overcome this technical difficulty in the field, several point-of-care devices have been developed in recent years for a rapid screening procedure [30,31,32]. These diagnostic devices characterize the G6PD phenotype of an individual by measuring enzyme activity and can be classified into qualitative and quantitative tests. The qualitative tests are designed to discriminate G6PD-deficient individuals from those with intermediate and normal enzyme activity, usually at the cut-off threshold of 30% of normal activity. They are particularly useful to screen G6PD deficiency in hemizygous males and homozygous females with two mutant alleles associated with G6PD deficiency [30, 31]. The design, items contained in the test packet, test cassette (or test strip), and procedures of qualitative tests for G6PD screening are similar to those of existing rapid diagnostic test (RDT) for malaria, which will facilitate the training of health personnel. A quantitative test may be more sensitive than qualitative tests to identify G6PD-deficient individuals and determine the level of activity in heterozygous females if its values are highly correlated with those of spectrophotometry. Although the results of both quantitative and qualitative tests can be obtained within 15 min, the principal disadvantages of a quantitative test are the higher cost and requirement for a specific instrument [30]. The present study was conducted with the aim to evaluate two rapid diagnostic test systems in Nouakchott: the qualitative CareStart™ RDT and the quantitative CareStart™ G6PD biosensor. Due to various financial and technical difficulties in obtaining a spectrophotometer and training a laboratory technician to use the equipment properly, G6PD genotyping was performed in the present study to serve as a gold standard against which the screening tools were measured.

Methods

Study sites and patients

Outpatients aged > 1 year old consulting spontaneously at Teyarett health centre, Sebkha health centre, or Maternity and Infant Hospital Centre (Centre Hospitalier mère-enfant) in Nouakchott were included in 2019–2020. The exclusion criteria were hospitalized patients, patients presenting at the emergency department, patients with severe disease, and patients with altered states of consciousness. After explaining the purpose of the study and obtaining an informed consent, basic information on age, sex, and ethnic origin was recorded in a survey questionnaire that was subsequently anonymised. Ethnic origin is a pertinent information in this study because Mauritania is a multi-ethnic country and ethnic origin is known to be associated with G6PD genotype [4]. No attempt was made in this study to include a representative and comparable sample of males and females of different ethnic origin.

Fingerprick capillary blood samples (approximately 200 µl) were collected to perform point-of-care rapid diagnostic tests for G6PD phenotype and determination of haemoglobin level. In addition, two or three drops of capillary blood (approximately 100–150 µl) were imbibed onto Whatman 3 mm filter paper, air dried, and stored at -20 °C for DNA analysis.

Phenotyping

G6PD phenotype was determined using two field-applicable point-of-care tools: the qualitative RDT for G6PD (CareStart™ G6PD RDT; Access Bio, Inc., Somerset, NJ) and quantitative CareStart™ G6PD biosensor (Access Bio, Inc., Somerset, NJ). The RDT for G6PD was performed to screen G6PD deficiency according to the manufacturer’s instructions. This RDT is designed to detect < 30% of normal G6PD activity, but it is most suitable for detecting class I and II G6PD deficiency [26], which corresponds to severe G6PD deficiency with < 10% of normal activity, according to the manufacturer. The quantitative G6PD biosensor was used in parallel to measure enzyme activity. Approximately 5 µl of capillary blood was transferred to the strip of the device. The biosensor determines G6PD enzyme activity in 4 min and automatically adjusts the reading to the ambient temperature. In addition to qualitative and quantitative tests for G6PD, haemoglobin level was measured using a portable HemoCue device (HemoCue AB, Ängelholm, Sweden) to normalize G6PD enzyme activity for haemoglobin.

Genotyping

Earlier studies have established that one of the most commonly observed G6PD genotypes in sub-Saharan Africa is the African-type G6PD A−, which is associated with a point mutation in nucleotide 376 and a second mutation in 202, 542, 680, or 968 [33,34,35,36]. In North Africa, as well as around the Mediterranean basin, the Mediterranean-type G6PD B−, which is associated with a single point mutation in nucleotide 563, is frequently found in people suffering from favism [37, 38]. Based on this information, nucleotide changes in 202, 376, 542, 563, 680, and 968 were investigated by polymerase chain reaction (PCR)-restriction fragment length polymorphism (RFLP) and/or sequencing in a stepwise manner in the present study.

In the first step, a fragment of exon 5 spanning nucleotide 376 and another fragment of exons 6–7 spanning nucleotide 563 were amplified separately by PCR using the protocol developed by Carter et al. [36] and Ezz El-Deen et al. [39]. All primer sequences were verified and corrected based on the complete G6PD sequence [2]. The fragments of exon 5 and exons 6–7 were digested by Fok I and Mbo II, respectively. The sample proceeded to the second step of the protocol if the mutant nucleotide 376G was found. PCR was performed to amplify a fragment of exon 4, which was digested with Nla III to determine nucleotide 202 (G6PD A−(202)) [36]. If the sample yielded 376G mutation in the first step and wild-type G202 in the second step, the third step consisted of PCR amplification of exon 9, followed by restriction digestion with Nci I [36]. If the wild-type T968 was found in the third step, exons 6–7 amplified in the initial step to determine nucleotide 563 were sequenced in the fourth step to determine nucleotides 542 and 680. Sequencing also allowed a second verification to exclude 563T (Mediterranean-type G6PD B−).

Data and statistical analysis

G6PD enzyme activity measured by CareStart G6PD biosensor was expressed as International Units/dL (IU/dL). G6PD activity was normalized for haemoglobin values measured at the same time as G6PD measurement and expressed as IU/g haemoglobin (IU/g Hb). G6PD activity is influenced by haemoglobin level. Anaemia, in particular, can increase G6PD activity due to reticulocytosis or on the contrary decrease G6PD activity due to chronic blood loss [26]. The WHO classification of haemoglobin levels to diagnose anaemia was used to classify patients who were anaemic at the time of G6PD measurement [40]. Based on this classification, non-anaemia, mild anaemia, moderate anaemia, and severe anaemia in male and female children < 5 years old were defined as ≥ 11.0 g/dL, 10.0–10.9 g/dL, 7.0–9.9 g/dL, and < 7.0 g/dL, respectively. In male and female children aged between 5 and 11 years old, non-anaemia, mild anaemia, moderate anaemia, and severe anaemia corresponded to ≥ 11.5 g/dL, 11.0–11.4 g/dL, 8.0–10.9 g/dL, and < 8.0 g/dL, respectively. In male and female children who were 12–14 years of age, as well as non-pregnant women aged ≥ 15 years old, non-anaemia, mild anaemia, moderate anaemia, and severe anaemia were defined as ≥ 12.0 g/dL, 11.0–11.9 g/dL, 8.0–10.9 g/dL, and < 8.0 g/dL, respectively. For men aged ≥ 15 years old, non-anaemia, mild anaemia, moderate anaemia, and severe anaemia were defined as ≥ 13.0 g/dL, 11.0–12.9 g/dL, 8.0–10.9 g/dL, and < 8.0 g/dL, respectively.

The genotype of normal G6PD enzyme without any mutations in males and females is designated GdB or, more commonly these days, G6PD B and BB, respectively [26]. If there was only one mutation in nucleotide 376, without any of the 4 nucleotide changes assessed in the present study, the genotype was referred to as G6PD A in hemizygous male and AA or BA in females. These three genotypes with a single A376G nucleotide change are not associated with G6PD deficiency. A normal G6PD without deficiency therefore refers to one of the following: A or B (in males), and AA, BB, or BA (in females). The African-type G6PD A− was defined as the presence of two mutations, A376G and one of the following: 376G + 202A (referred to as G6PD A−(202)), 376G + 968C (called G6PD A−(968) or Betica-Selma), 376G + 542T (called G6PD A− Santamaria), or 376G + 680T (G6PD A−(680)). The African-type genotype is designated A− in hemizygous male, A−A− in homozygous female, and AA− or BA− in heterozygous female. In the presence of 563T mutation (without the African-type G6PD mutations since they are mutually exclusive), the sample was classified as the Mediterranean-type G6PD B−.

Descriptive statistics and performance of the screening tests were calculated using GraphPad Prism (GraphPad software; San Diego, CA). Qualitative variables were compared using Fisher’s exact test. The level of significance of all statistical tests was set at P < 0.05.

Results

Characteristics of patients

A total of 323 patients (74 males and 249 females) were included in the present study (Table 1). The sex ratio (M:F, 0.30) was highly skewed towards adult women because most patients (n = 255) were recruited at the Maternity and Infant Hospital Centre. A total of 68 patients were recruited at the other health centres (23 at Teyarett health centre and 45 at Sebkha health centre). The overall mean age [± standard deviation (SD); range] of the patients was 23.6 ± 14.4 (1.1–91) years old. The mean age of males (11.4 years old) was significantly younger (P < 0.05) than that of females (27.2 years old). There were 142 (44.0%) white Moors, 74 (22.9%) black Moors, 67 (20.7%) Pulars, 23 (7.1%) Wolofs, 10 (3.1%) Soninkes, 1 (0.3%) black Moor-Pular, 1 (0.3%) black Moor-Wolof, 1 (0.3%) Pular-Soninke, and 4 (1.2%) patients of unknown ethnic origin (missing data). The mean haemoglobin (± SD; range) of the patients was 11.3 ± 1.8 g/dL (5.5–16.7 g/dL). A total of 37 (50.0%; 95% CI: 38.1–61.9%) males and 141 (56.6%; 95% CI: 50.2–62.9%) females were anaemic, based on sex- and age-adjusted cut-off thresholds established by the WHO (Table 1; Fig. 1) [40]. The majority of these patients (n = 145) had mild to moderate level of anaemia; 24 had severe anaemia, mostly near the borderline level.

Table 1 Characteristics of included patients
Fig. 1
figure 1

Distribution of individual haemoglobin values of Mauritanian patients included in the study according to sex (males, blue; females, red) and age groups (years, yr) used in the WHO classification [40]. The horizontal bars denote cut-offs above which the patient is considered non-anaemic

G6PD genotypes

Of 74 males, 69 had either the wild-type genotype B (n = 61) or genotype A (i.e., with a single A376G mutation; n = 8), both of which are associated with normal G6PD activity (Table 2). Five carried the African-type A- genotype: G6PD A−(202) G202A/A376G (n = 3) or G6PD A−(968) Betica-Selma A376G/T968C (n = 2) (Table 3). Among 5 males with the African-type A− genotype, 3 were black Moors (2 G6PD A−(202) and 1 G6PD A−(968)), 1 was white Moor (G6PD A−(202)), and 1 was Soninke (G6PD A−(968)).

Table 2 G6PD genotypes of Mauritanian patients included in the study
Table 3 Frequency of G6PD A− allelic variants in individuals with G6PD A− genotypes in Nouakchott

Among females, 219 had genotypes associated with normal G6PD (177 BB, 37 BA, and 5 AA), 28 were heterozygotes (5 AA− and 23 BA−), and 2 were homozygotes (A−A−) (Table 2). Among heterozygotes, 13 carried G6PD A−(202), 12 G6PD A−(968) Betica-Selma, and 3 G6PD A−(542) Santamaria variants (Table 3). Heterozygous females belonged to white Moors (n = 4) and Mauritanians of black African ancestry (black Moors, n = 10; Pular, n = 12; Wolof, n = 1; Soninke, n = 1). Both homozygous women (one black Moor and one Soninke) had the African-type A−(202) (376G/202A). None of the patients of both sexes had the Mediterranean-type G6PD genotype.

Screening tests

The distribution of G6PD activity determined by CareStart G6PD biosensor in Mauritanians included in the present study is illustrated in Fig. 2. Overall, the mean [± SD; 95% confidence intervals (95% CI)] activity was 5.35 ± 3.24 (4.94–5.70) IU/g Hb in 323 patients. The median (range) was 5.19 (0.08–21.94) IU/g Hb. The mean (± SD; 95% CI) and median (range) activities in different subsets of patient population were calculated (Table 4). The median value in the male population (n = 72) was 5.92 IU/g Hb. The adjusted median value in the male population, calculated by excluding those with genotypes associated with G6PD deficiency (i.e., hemizygous A−), was 6.18. The distribution of anaemia in different age groups in both sexes (Fig. 1) and the distribution of G6PD enzyme activity according to sex and age groups (Fig. 3) suggested that bias may have been introduced due to the presence of many patients with different levels of anaemia. To correct this bias, further calculations were made to reduce the effect of anaemia on G6PD activity (Table 4). In non-anaemic male population (n = 35) and male and female population (n = 130) with normal genotypes (i.e., A or B in males and AA, BB, or BA in females), the median G6PD activities were 5.72 and 5.34 IU/g Hb, respectively. Although the adjusted median of the male population (i.e., 6.18 U/g Hb) would have been used to set the 100% activity in a study using spectrophotometry as the reference method, the adjusted median of 5.34 IU/g Hb, determined from non-anaemic male and female population, was considered to correspond to 100% G6PD activity in the present study, in which genotyping was the reference method.

Fig. 2
figure 2

Glucose-6-phosphate dehydrogenase (G6PD) activity normalized for haemoglobin [International Unit/g haemoglobin (IU/g Hb)] was determined in Mauritanian patients (n = 323) by CareStart G6PD biosensor in Nouakchott

Table 4 Profile of G6PD activity in Mauritanian patients
Fig. 3
figure 3

The dashed line denotes the median value of glucose-6-phosphate dehydrogenase (G6PD) activity found in male and female non-anaemic children aged > 1 year old (year) and adults (5.34 International Units/g haemoglobin [IU/g Hb]). One outlier datapoint (21.9 IU/g Hb) found in a girl aged less than 5 years is not presented in the graph. The age groups are those used in the World Health Organization (WHO) classification of haemoglobin levels to diagnose anaemia [40]

CareStart G6PD RDT was positive in 8 patients. Among seven patients with A− or A−A− genotype, all but one were correctly identified as G6PD deficient by RDT. Six of these patients had G6PD activity ranging from 0.12 to 1.43 IU/g Hb, as measured by biosensor. One homozygous female with A−(202) genotype in whom RDT yielded a negative result had G6PD activity of 2.34 IU/g Hb.

Performance of G6PD biosensor

The G6PD activities in three males with G6PD A−(202) genotype were 0.29 IU/g Hb (Hb 10.3 g/dL), 0.59 IU/g Hb (Hb 10.1 g/dL), and 1.43 IU/g Hb (Hb 13.3 g/dL) and those in two males with G6PD A−(968) Betica-Selma were 0.12 IU/g Hb (Hb 8.2 g/dL) and 0.38 IU/g Hb (Hb 8.0 g/dL). Based on 100% activity set at 5.34 IU/g Hb, these activities in males with A− genotypes corresponded to 2.2–26.8% of the normal activity. CareStart G6PD RDT was positive in all of these five male patients, indicating G6PD deficiency. Two homozygous females with A−(202) genotype had G6PD activities of 2.34 IU/g Hb (Hb, 10.7 g/dL; RDT negative) and 0.14 IU/g Hb (Hb, 7.3 g/dL; RDT positive), corresponding to 43.8% and 2.6% of the normal enzyme activity, respectively.

The performance of G6PD biosensor is best assessed against spectrophotometric measurement of G6PD activity not only because the latter is the gold standard but also because they are both phenotypic tests. Nonetheless, in the present study, the performance of G6PD biosensor was assessed against genotyping (Table 5). AA− (n = 5) and BA− (n = 23) heterozygous females, but not BA heterozygotes, were excluded from analysis because it is known that their phenotype cannot be predicted from genotype. At 10% cut-off level, its sensitivity was low (57.1%), and the specificity was high (98.6%). At 30% cut-off, the sensitivity and specificity were 85.7% and 91.7%, respectively. At 80% cut-off, the sensitivity was 100% while the specificity was 64.9%. The sensitivity and the specificity did not improve when patients with severe anaemia (n = 24) were excluded from analysis. Further analysis in a subset of patients without mild to moderate anaemia was not performed because only one patient with the G6PD genotype associated with African-type G6PD deficiency was non-anaemic.

Table 5 Performance of CareStart G6PD biosensor against genotyping

Assuming the reliability of G6PD biosensor measurement, the possibility to administer primaquine to patients on the basis of G6PD activity was simulated and assessed (note: the patients included in the present study were not malaria-infected). The results are summarized in Table 6. About one-fourth of males and females included in this study had at least 80% of G6PD activity, which was supported by genotyping showing one of the genotypes associated with normal G6PD activity. The presence of > 80% of G6PD activity is compatible with the safe administration of a standard dose of primaquine (0.25 mg base/kg body weight/day for 14 days) or tafenoquine (300 mg single dose) to kill hypnozoites [21, 22, 27]. About 15% and 3.4% of males and females of the study population had 30–80% and < 30% of residual enzyme activity, respectively. These patients would have been eligible for a modified primaquine treatment regimen, which consists of a weekly dose (0.75 mg base) for 8 weeks. Six of seven hemizygous and homozygous patients with the African-type genotype were anaemic, and four of these six anaemic patients had enzyme activity of 0.12–0.38 IU/g Hb, i.e., 2.2–7.1% of activity. The other two anaemic patients with African-type genotypes had 0.59 IU/g Hb (11.0% activity) and 2.34 IU/g Hb (43.8% activity). More than half of the patient population (54–57%) in the present study had varying degrees of anaemia, which may lead to an unreliable measurement of G6PD activity. In real situations, before primaquine treatment can be re-considered for these patients, quantitative G6PD testing will be required after anaemia is corrected.

Table 6 G6PD activity level and possibility to administer primaquine

Discussion

As one of the first steps to facilitate implementation of the new drug policy including primaquine, the current prevalence of G6PD genotypes in Mauritania was assessed in our recent works [41, 42]. The preliminary database showed that, overall, 6.3% of Mauritanian men and 1.5% of Mauritanian women are affected by G6PD genetic polymorphisms leading to the African-type G6PD deficiency, and that the Mediterranean-type G6PD deficiency is absent [42]. Moreover, these studies have shown that G6PD deficiency tends to occur much more frequently in Mauritanian ethnic groups of Black African ancestry. The present study is the second phase of the preliminary efforts to control P. vivax malaria in the country. CareStart™ G6PD RDT and CareStart™ G6PD biosensor were assessed under real conditions in health centres and in a hospital in Nouakchott.

CareStart™ G6PD RDT performed poorly in the present study, contrary to the conclusion of systematic reviews and meta-analysis of earlier studies conducted elsewhere [28, 31]. Although the RDT was positive, indicating < 30% of G6PD activity, in six of seven patients with African-type genotype associated with G6PD deficiency, numerous false negatives were observed. One possible explanation may be poor storage condition and/or high ambient temperature during screening. The recommended storage condition of the kit is ambient temperature, and the optimal temperature for its use is between 18 and 32 °C. The ambient temperature can exceed 35 °C in Nouakchott during certain months of the year. Another drawback of this qualitative screening test is the difficulty of interpretation when there is a slight colour change. The difference between normal G6PD activity (distinct colour change to violet) and deficiency (slight colour change to violet or no colour change) may be difficult to distinguish, especially because there is no control band that indicates that the test is valid or not. Further improvements are needed to conserve the quality of RDT for G6PD in a hot, dry climate, as in most of Mauritania where P. vivax occurs.

By contrast, CareStart™ G6PD biosensor is a user-friendly point-of-care device with automatic temperature adjustment that performed better than CareStart™ G6PD RDT when compared to G6PD genotyping at the cut-off level of 80%. The first-generation G6PD biosensor was used in the present study. Other earlier studies also reported that the performance of CareStart™ G6PD biosensor needs further improvement. In a study conducted by Weppelmann et al. in Haiti [43], the sensitivity and the specificity of biosensor, compared to spectrophotometry, were 53.7% and 94.6% at < 60% G6PD activity and 5.9% and 99.7% at < 30% activity, respectively. Alam et al. [44] found the sensitivity and the specificity of biosensor, compared to spectrophotometry, of 71% and 98% at < 70% enzyme activity, the level beyond which tafenoquine can be administered, in Bangladesh. The optimal cut-off to obtain 91% sensitivity and 82% specificity at < 70% enzyme activity was 6.8 IU/g Hb after adjustment of 100% activity. The sensitivity and the specificity of the second-generation biosensor, compared to spectrophotometry, were similar, i.e., 79% and 98%, respectively, at 70% cut off in another study performed in Thai-Burmese border only after the initial threshold for biosensor was corrected using Youden’s index [45]. In our study, the adjustment of median G6PD values using receiver operating characteristic (ROC) curve analysis and Youden’s index to ameliorate the sensitivity and the specificity of G6PD biosensor was not attempted as in other studies because spectrophotometry was not used to validate the measurements of biosensor [44,45,46].

The results of G6PD biosensor in our study generally support those of other authors conducted in different epidemiological context [43,44,45,46]. In practice, however, there are several caveats associated with the use of this device. First, haemoglobin measurement is required in parallel with G6PD activity. The first-generation CareStart G6PD biosensor used in the present study does not have a measurement of haemoglobin incorporated in the device, necessitating a separate device to determine the haemoglobin level. In patients who are anaemic at the time of G6PD testing, the result of G6PD activity may be either normal or high due to reticulocytosis or low due to blood loss. In either case, G6PD activity during the anaemic phase does not reflect the baseline G6PD activity of the individual. If known, the cause(s) of anemia should be treated first, and G6PD testing repeated after several weeks. The drawback is that there is currently no clear guideline on whether to or how to administer anti-hypnozoite drug in such circumstances. Second, before primaquine is administered to non-anaemic women, a precise measurement of G6PD enzyme activity with a reliable quantitative assay is required, as qualitative screening tests cannot reliably detect G6PD deficiency in heterozygous women. In this context, the enzymatic activity measured by G6PD biosensor may have to be calibrated against spectrophotometry for each patient population in different countries. Until present, spectrophotometric measurement of G6PD activity has not been performed in Mauritania. G6PD genotyping may circumvent this problem, but it is inconceivable to extend this practice in the field. Third, there is no universally accepted normal range for G6PD activity. Each country is required to determine the normal G6PD activity of its population measured with a quantitative gold standard procedure. The cut-off value calculated in the present study cannot be extrapolated to other methods or to populations who were excluded, in particular new-borns and infants less than 1 year old. Fourth, even if anaemic patients are excluded from immediate treatment with primaquine, other haematological conditions, such as haemoglobinopathies and leukocytaemia, may yield a normal G6PD test result in a G6PD-deficient patient [6].

One of the limitations of the present study is the lack of comparison of the performance of biosensor with that of the gold standard, spectrophotometry. Although other authors have confirmed the high reliability of biosensor, which gave results that are highly correlated with those of spectrophotometry [43,44,45,46,47], high correlation does not imply that biosensor yields the same enzyme activity as spectrophotometry. In the study conducted by Alam et al. [44], the correlation of enzyme activity measurements between biosensor and spectrophotometry was 0.773, and discordant results were obtained in some patient samples at < 30% enzyme level. Bancone et al. also reported that CareStart G6PD biosensor overestimates enzyme activity compared to spectrophotometry [45]. As a pilot study to assess the feasibility and practicality of CareStart G6PD biosensor, the present study was not designed to determine 100% G6PD activity and cut-off levels. The value of 100% G6PD activity based on the adjusted median found in the present study should therefore be considered as a preliminary estimate in Mauritania. Future work plan before implementing P. vivax elimination programme should include a comparison of G6PD activity measured by biosensor and spectrophotometry in a large sample of general population representing different ethnic backgrounds with an aim to define the normal values in the Mauritanian population. Another limitation of this study is the overrepresentation of females in our study population. For studies on an X-linked disorder, such as G6PD deficiency, studies in males would yield a higher number of subjects with the disorder to facilitate the calculation of cut-off threshold of G6PD deficiency while limiting the major confounding factor, i.e., heterozygosity associated with unpredictable G6PD activity in females due to the phenomenon called lyonisation [48,49,50]. However, other studies on the assessment of G6PD biosensor have also included more females than males, as in the works of Alam et al. (73% females) and Bancone et al. (72% females) [44, 45]. Genotyping is at present the only method that allows an accurate identification of heterozygous females. More investigations on G6PD involving females are required to resolve the problem of primaquine treatment in heterozygous females. Moreover, an updated version of WHO classification of G6PD variants and recommendation for 8-aminoquinoline treatment is needed for implementation of P. vivax control programmes in Mauritania and elsewhere in the world [51].

Conclusions

The present study is the first of its kind in Mauritania that assessed the utility of G6PD biosensor. Because of the predominance of P. vivax malaria in Nouakchott and elsewhere in northern part of the country, a point-of-care G6PD testing is of primordial importance in ensuring that primaquine is administered to P. vivax-infected patients with normal G6PD activity and that treatment leads to complete cure. Further work with G6PD biosensor, in parallel with spectrophotometric measurement, will be required before full deployment of this point-of-care tool in the field in Mauritania.

Availability of data and materials

All data generated and analysed during this study are included in this published article.

Abbreviations

G6PD:

Glucose-6-phosphate dehydrogenase

NADPH:

Nicotinamide adenine dinucleotide phosphate hydrogen

PCR:

Polymerase chain reaction

RDT:

Rapid diagnostic test

ROC:

Receiver operating characteristic

RFLP:

Restriction fragment length polymorphism

UV:

Ultraviolet

WHO:

World Health Organization

References

  1. Takizawa T, Huang IY, Ikuta T, Yoshida A. Human glucose-6-phosphate dehydrogenase: primary structure and cDNA cloning. Proc Natl Acad Sci USA. 1986;83(12):4157–61.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Chen EY, Cheng A, Lee A, Kuang WJ, Hillier L, Green P, et al. Sequence of human glucose-6-phosphate dehydrogenase cloned in plasmids and a yeast artificial chromosome. Genomics. 1991;10(3):792–800.

    Article  CAS  PubMed  Google Scholar 

  3. Yoshida A, Beutler E, Motulsky AG. Human glucose-6-phosphate dehydrogenase variants. Bull World Health Organ. 1971;45(2):243–53.

    CAS  PubMed  PubMed Central  Google Scholar 

  4. Minucci A, Moradkhani K, Hwang MJ, Zuppi C, Giardina B, Capoluongo E. Glucose-6-phosphate dehydrogenase (G6PD) mutations database: review of the “old” and update of the new mutations. Blood Cells Mol Dis. 2012;48(3):154–65.

    Article  CAS  PubMed  Google Scholar 

  5. Gómez Manzo S, Marcial Quino J, Vanoye Carlo A, Serrano Posada H, Ortega Cuellar D, González Valdez A, et al. Glucose-6-phosphate dehydrogenase: update and analysis of new mutations around the world. Int J Mol Sci. 2016;17(12):2069.

    Article  PubMed Central  CAS  Google Scholar 

  6. World Health Organization Working Group. Glucose-6-phosphate dehydrogenase deficiency. Bull World Health Organ. 1989;67(6):601–11.

    Google Scholar 

  7. Howes RE, Battle KE, Satyagraha AW, Baird JK, Hay SI. G6PD deficiency: global distribution, genetic variants and primaquine therapy. Adv Parasitol. 2013;81:133–201.

    Article  PubMed  Google Scholar 

  8. Tarlov AR, Brewer GJ, Carson PE, Alving AS. Primaquine sensitivity: glucose-6-phosphate dehydrogenase deficiency: an inborn error of metabolism of medical and biological significance. Arch Intern Med. 1962;109:209–34.

    Article  CAS  PubMed  Google Scholar 

  9. Degowin RL, Eppes RB, Powell RD, Carson PE. The haemolytic effects of diaphenylsulfone (DDS) in normal subjects and in those with glucose-6-phosphate-dehydrogenase deficiency. Bull World Health Organ. 1966;35(2):165–79.

    CAS  PubMed  PubMed Central  Google Scholar 

  10. Shanks GD. Drugs for prophylaxis and treatment of malaria. J Travel Med. 1994;1(1):40–7.

    Article  CAS  PubMed  Google Scholar 

  11. Luzzatto L. The rise and fall of the antimalarial Lapdap: a lesson in pharmacogenetics. Lancet. 2010;376(9742):739–41.

    Article  PubMed  Google Scholar 

  12. Pamba A, Richardson ND, Carter N, Duparc S, Premji Z, Tiono AB, et al. Clinical spectrum and severity of hemolytic anemia in glucose 6-phosphate dehydrogenase-deficient children receiving dapsone. Blood. 2012;120(20):4123–33.

    Article  CAS  PubMed  Google Scholar 

  13. Müller O, Mockenhaupt FP, Marks B, Meissner P, Coulibaly B, Kuhnert R, et al. Haemolysis risk in methylene blue treatment of G6PD-sufficient and G6PD-deficient West-African children with uncomplicated falciparum malaria: a synopsis of four RCTs. Pharmacoepidemiol Drug Saf. 2013;22(4):376–85.

    Article  PubMed  CAS  Google Scholar 

  14. Hounkpatin AB, Kreidenweiss A, Held J. Clinical utility of tafenoquine in the prevention of relapse of Plasmodium vivax malaria: a review on the mode of action and emerging trial data. Infect Drug Resist. 2019;12:553–70.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Beutler E. The hemolytic effect of primaquine and related compounds: a review. Blood. 1959;14(2):103–39.

    Article  CAS  PubMed  Google Scholar 

  16. WHO. Guidelines for the treatment of malaria. Third edition. Geneva, Switzerland: World Health Organization Publications; 2015. https://www.afro.who.int/publications/guidelines-treatment-malaria-third-edition (accessed 7 July 2021).

  17. Recht J, Ashley E, White N. Safety of 8-aminoquinoline antimalarial medicines. Geneva, Switzerland: World Health Organization Publications; 2014. https://apps.who.int/iris/handle/10665/112735 (accessed 7 July 2021).

  18. WHO. Policy brief on single-dose primaquine as a gametocytocide in Plasmodium falciparum malaria. WHO/HTM/GMP/2015.1; Geneva, Switzerland; World Health Organization Publication; 2015. https://apps.who.int/iris/handle/10665/338498 (accessed 7 July 2021).

  19. Graves PM, Choi L, Gelband H, Garner P. Primaquine or other 8-aminoquinolines for reducing Plasmodium falciparum transmission. Cochrane Database Syst Rev. 2018;2(2):CD008152.

    PubMed  Google Scholar 

  20. Chu CS, White NJ. Management of relapsing Plasmodium vivax malaria. Expert Rev Anti Infect Ther. 2016;14(10):885–900.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Lacerda MVG, Llanos-Cuentas A, Krudsood S, Lon C, Saunders DL, Mohammed R, et al. Single-dose tafenoquine to prevent relapse of Plasmodium vivax malaria. N Engl J Med. 2019;380(3):215–28.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Llanos-Cuentas A, Lacerda MVG, Hien TT, Vélez ID, Namaik-Larp C, Chu CS, et al. Tafenoquine versus primaquine to prevent relapse of Plasmodium vivax malaria. N Engl J Med. 2019;380(3):229–41.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Lekweiry KM, Salem MS, Basco LK, Briolant S, Hafid J, Boukhary AO. Malaria in Mauritania: retrospective and prospective overview. Malar J. 2015;14:100.

    Article  PubMed  PubMed Central  Google Scholar 

  24. WHO. Guidelines for the treatment of malaria. 2nd edition. Geneva, Switzerland: World Health Organization Publication; 2010. https://www.paho.org/en/node/50095 (accessed 7 July 2021).

  25. Mauritanian Ministry of Health. Guide clinique et thérapeutique à l’usage du personnel des centres de santé de la Mauritanie. 2nd ed. Madrid: Agencia Española de Cooperación Internacional para el Desarrollo (AECID); 2013. http://www.sante.gov.mr/?wpfb_dl=140 (accessed 9 April 2021).

  26. WHO. Standardization of procedures for the study of glucose-6-phosphate dehydrogenase. WHO Tech Rep Ser 366. Geneva, Switzerland: World Health Organization Publication; 1967. https://apps.who.int/iris/handle/10665/40660 (accessed 7 July 2021).

  27. Domingo GJ, Satyagraha AW, Anvikar A, Baird K, Bancone G, Bansil P, et al. G6PD testing in support of treatment and elimination of malaria: recommendations for evaluation of G6PD tests. Malar J. 2013;12:391.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  28. Ley B, Bancone G, von Seidlein L, Thriemer K, Richards JS, Domingo GJ, et al. Methods for the field evaluation of quantitative G6PD diagnostics: a review. Malar J. 2017;16(1):361.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  29. Pfeffer DA, Ley B, Howes RE, Adu P, Alam MS, Bansil P, et al. Quantification of glucose-6-phosphate dehydrogenase activity by spectrophotometry: a systematic review and meta-analysis. PLoS Med. 2020;17(5):e1003084.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Anderle A, Bancone G, Domingo GJ, Gerth-Guyette E, Pal S, Satyagraha AW. Point-of-care testing for G6PD deficiency: opportunities for screening. Int J Neonatal Screen. 2018;4(4):34.

    Article  PubMed  PubMed Central  Google Scholar 

  31. WHO. Malaria Policy Advisory Committee Meeting. Point-of-care G6PD testing to support safe use of primaquine for the treatment of vivax malaria. WHO Evidence Review Group meeting report, 8–9 October 2014; Geneva, Switzerland, 2015; WHO/HTM/GMP/MPAC/2015.6. http://www.who.int/malaria/mpac/mpac-mar2015-erg-g6pd-presentation.pdf (accessed 7 July 2021).

  32. Grobusch MP, Rodríguez-Morales AJ, Schlagenhauf P. The primaquine problem-and the solution? Point-of-care diagnostics for glucose-6-phosphate dehydrogenase deficiency. Clin Infect Dis. 2019;69(8):1443–5.

    Article  CAS  PubMed  Google Scholar 

  33. Takizawa T, Yoneyama Y, Miwa S, Yoshida A. A single nucleotide base transition is the basis of the common human glucose-6-phosphate dehydrogenase variant A (+). Genomics. 1987;1(3):228–31.

    Article  CAS  PubMed  Google Scholar 

  34. Hirono A, Beutler E. Molecular cloning and nucleotide sequence of cDNA for human glucose-6-phosphate dehydrogenase variant A(-). Proc Natl Acad Sci USA. 1988;85(11):3951–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Beutler E, Kuhl W, Vives-Corrons JL, Prchal JT. Molecular heterogeneity of glucose-6-phosphate dehydrogenase A-. Blood. 1989;74(7):2550–5.

    Article  CAS  PubMed  Google Scholar 

  36. Carter N, Pamba A, Duparc S, Waitumbi JN. Frequency of glucose-6-phosphate dehydrogenase deficiency in malaria patients from six African countries enrolled in two randomized anti-malarial clinical trials. Malar J. 2011;10:241.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  37. Vulliamy TJ, D’Urso M, Battistuzzi G, Estrada M, Foulkes NS, Martini G, et al. Diverse point mutations in the human glucose-6-phosphate dehydrogenase gene cause enzyme deficiency and mild or severe hemolytic anemia. Proc Natl Acad Sci USA. 1988;85(14):5171–5.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  38. Vives-Corrons JL, Kuhl W, Pujades MA, Beutler E. Molecular genetics of the glucose-6-phosphate dehydrogenase (G6PD) Mediterranean variant and description of a new G6PD mutant, G6PD Andalus1361A. Am J Hum Genet. 1990;47(3):575–9.

    CAS  PubMed  PubMed Central  Google Scholar 

  39. Ezz El-Deen ZM, Hussin NF, Abdel Hamid TA, Abdel Migeed OR, Samy RM. G6PD deficiency and G6PD (Mediterranean and silent) polymorphisms in Egyptian infants with neonatal hyperbilirubinemia. Lab Med. 2013;44(3):228–34.

    Article  Google Scholar 

  40. WHO. Haemoglobin concentrations for the diagnosis of anaemia and assessment of severity. WHO/NMH/NHD/MNM/11.1, Vitamin and Mineral Nutrition Information System. Geneva, Switzerland: World Health Organization Publication; 2011. (http://www.who.int/vmnis/indicators/haemoglobin.pdf) (accessed 7 July 2021).

  41. Djigo OKM, Bollahi MA, Hasni Ebou M, Ould Ahmedou Salem MS, Tahar R, Bogreau H, et al. Assessment of glucose-6-phosphate dehydrogenase activity using CareStart G6PD rapid diagnostic test and associated genetic variants in Plasmodium vivax malaria endemic setting in Mauritania. PLoS ONE. 2019;14(9):0220977.

    Article  CAS  Google Scholar 

  42. Djigo OKM, Ould Ahmedou Salem MS, Diallo SM, Bollahi MA, Boushab MB, Garre A, et al. Molecular epidemiology of G6PD genotypes in different ethnic groups residing in Saharan and Sahelian zones of Mauritania. Pathogens. 2021;10:931.

  43. Weppelmann TA, von Fricken ME, Wilfong TD, Aguenza E, Philippe TT, Okech BA. Field trial of the CareStart biosensor analyzer for the determination of glucose-6-phosphate dehydrogenase activity in Haiti. Am J Trop Med Hyg. 2017;97(4):1262–70.

    Article  CAS  PubMed  Google Scholar 

  44. Alam MS, Kibria MG, Jahan N, Thriemer K, Hossain MS, Douglas NM, et al. Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity. PLoS ONE. 2018;13(11):0206331.

    Google Scholar 

  45. Bancone G, Gornsawun G, Chu CS, Porn P, Pal S, Bansil P, et al. Validation of the quantitative point-of-care CareStart biosensor for assessment of G6PD activity in venous blood. PLoS ONE. 2018;13(5):0196716.

    Article  CAS  Google Scholar 

  46. Pal S, Bansil P, Bancone G, Hrutkay S, Kahn M, Gornsawun G, et al. Evaluation of a novel quantitative test for glucose-6-phosphate dehydrogenase deficiency: bringing quantitative testing for glucose-6-phosphate dehydrogenase deficiency closer to the patient. Am J Trop Med Hyg. 2019;100(1):213–21.

    Article  CAS  PubMed  Google Scholar 

  47. Pengboon P, Thamwarokun A, Changsri K, Kaset C, Chomean S. Evaluation of quantitative biosensor for glucose-6-phosphate dehydrogenase activity detection. PLoS ONE. 2019;14(12):0226927.

    Article  CAS  Google Scholar 

  48. Lyon MF. Gene action in the X-chromosome of the mouse (Mus musculus L.). Nature. 1961;190:372–3.

    Article  CAS  PubMed  Google Scholar 

  49. Harper PS. Mary Lyon and the hypothesis of random X chromosome inactivation. Hum Genet. 2011;130(2):169–74.

    Article  CAS  PubMed  Google Scholar 

  50. Shvetsova E, Sofronova A, Monajemi R, Gagalova K, Draisma HHM, White SJ, et al. Skewed X-inactivation is common in the general female population. Eur J Hum Genet. 2019;27(3):455–65.

    Article  CAS  PubMed  Google Scholar 

  51. WHO Malaria Policy Advisory Committee. Updating the WHO G6PD classification of variants and the International Classification of Diseases, 11th revision (ICD-11). WHO/CDS/GMP/MPAC/2019.15. Geneva, Switzerland: World Health Organization Publications; 2019. https://www.who.int/malaria/mpac/mpac-october2019-session7-updating-G6PD-classification.pdf (accessed 7 July 2021).

Download references

Acknowledgements

We are grateful to all patients and parents and/or legal guardians of children who accepted to participate in this study.

Funding

This study was supported by Institut de Recherche pour le Développement (IRD), Jeune Equipe Associée à l’IRD (RI3M), Institut de Recherche Biomédicale des Armées (IRBA), and Direction Générale de l’Armement (DGA; Grant number PDH-2-NRBC-2-B-2113). O.K.M.D. received a PhD scholarship from the French government.

Author information

Authors and Affiliations

Authors

Contributions

OKMD and MSOAS collected blood samples and performed G6PD biosensor. YOK, MSOAS, and AOMSB supervised the study in health centres and hospital. OKMD, MSOAS, LB and SB developed the database. OKMD and NG performed PCR. LB and SB designed the PCR protocol. LB and SB analysed and interpreted clinical and genotyping data. LB wrote the draft. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ali Ould Mohamed Salem Boukhary.

Ethics declarations

Ethics approval and consent to participate

The purpose of the study was explained in local dialect to all adult patients, to the parents or legal guardians of paediatric patients, or to adolescents aged between 12 and 18 years old. Written informed consent was obtained from all adult participants, or from the parents or legal guardians accompanying children aged < 12 years old, before collection of blood samples. For adolescents aged 12–18 years old, their assent was also obtained in addition to their parents’ or legal guardian’s consent. The study protocol was reviewed and approved by the institutional ethics committee of the University of Nouakchott Al-Aasriya, Nouakchott, Mauritania (approval no. 112/12-09-2014/USTM, 003/2020/CE/UNA) and the institutional review board of the Institut de Recherche pour le Développement (IRD), Marseille, France (Comité consultative de déontologie et d’éthique approval no. 15/12/2012).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Djigo, O.K.M., Ould Khalef, Y., Ould Ahmedou Salem, M.S. et al. Assessment of CareStart G6PD rapid diagnostic test and CareStart G6PD biosensor in Mauritania. Infect Dis Poverty 10, 105 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s40249-021-00889-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s40249-021-00889-2

Keywords